Basit öğe kaydını göster

dc.contributor.authorDemirci, Utkanen_US
dc.contributor.authorErgün, Arif Sanlıen_US
dc.contributor.authorOralkan, Ömeren_US
dc.contributor.authorKaraman, Mustafaen_US
dc.contributor.authorKhuri-Yakub, Butrus Thomasen_US
dc.date.accessioned2015-01-15T23:00:08Z
dc.date.available2015-01-15T23:00:08Z
dc.date.issued2004-07
dc.identifier.citationDemirci, U., Ergün, A. S., Oralkan, Ö. & Khuri-Yakub, B. T. (2004). Forward-viewing CMUT arrays for medical imaging. IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control, 49; 51(7), 887-895. doi:10.1109/TUFFC.2004.1320749en_US
dc.identifier.issn0885-3010
dc.identifier.issn1525-8955
dc.identifier.urihttps://hdl.handle.net/11729/154
dc.identifier.urihttp://dx.doi.org/10.1109/TUFFC.2004.1320749
dc.description.abstractThis paper reports the design and testing of forward-viewing annular arrays fabricated using capacitive micromachined ultrasonic transducer (CMUT) technology. Recent research studies have shown that CMUTs have broad frequency bandwidth and high-transduction efficiency. One- and two-dimensional CMUT arrays of various sizes already have been fabricated, and their viability for medical imaging applications has been demonstrated. We fabricated 64-element, forward-viewing annular arrays using the standard CMUT fabrication process and carried out experiments to measure the operating frequency, bandwidth, and transmit/receive efficiency of the array elements. The annular array elements, designed for imaging applications in the 20 MHz range, had a resonance frequency of 13.5 MHz in air. The immersion pulse-echo data collected from a plane reflector showed that the devices operate in the 5-26 MHz range with a fractional bandwidth of 135%. The output pressure at the surface of the transducer was measured to be 24 kPa/V. These values translate into a dynamic range of 131.5 dB for I-V excitation in 1-Hz bandwidth with a commercial low noise receiving circuitry. The designed, forward-viewing annular CMUT array is suitable for mounting on the front surface of a cylindrical catheter probe and can provide Doppler information for measurement of blood flow and guiding information for navigation through blood vessels in intravascular ultrasound imaging.en_US
dc.language.isoengen_US
dc.publisherIEEE-INST Electrical Electronics Engineers Incen_US
dc.relation.isversionof10.1109/TUFFC.2004.1320749
dc.rightsinfo:eu-repo/semantics/closedAccessen_US
dc.subjectMicromachined ultrasonic transducersen_US
dc.subjectCapacitance transducersen_US
dc.subjectImmersion transducersen_US
dc.subjectTransducersen_US
dc.subjectElectric capacitanceen_US
dc.subjectEquipment designen_US
dc.subjectEquipment failure analysisen_US
dc.subjectMiniaturizationen_US
dc.subjectReproducibility of resultsen_US
dc.subjectSensitivity and specificityen_US
dc.subjectUltrasonographyen_US
dc.subjectArraysen_US
dc.subjectBandwidthen_US
dc.subjectCapacitanceen_US
dc.subjectDoppler effecten_US
dc.subjectFrequenciesen_US
dc.subjectMicromachiningen_US
dc.subjectProduct designen_US
dc.subjectResonanceen_US
dc.subjectSpurious signal noiseen_US
dc.subjectUltrasonic transducersen_US
dc.subjectAnnular arraysen_US
dc.subjectOutput pressureen_US
dc.subjectPulse echoen_US
dc.subjectArticleen_US
dc.subjectEchographyen_US
dc.subjectElectronicsen_US
dc.subjectEquipmenten_US
dc.subjectEquipment designen_US
dc.subjectEvaluationen_US
dc.subjectInstrumentationen_US
dc.subjectMethodologyen_US
dc.subjectReproducibilityen_US
dc.subjectMedical imagingen_US
dc.titleForward-viewing CMUT arrays for medical Imagingen_US
dc.typearticleen_US
dc.description.versionPublisher's Versionen_US
dc.relation.journalIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Controlen_US
dc.contributor.departmentIşık Üniversitesi, Mühendislik Fakültesi, Elektrik-Elektronik Mühendisliği Bölümüen_US
dc.contributor.departmentIşık University, Faculty of Engineering, Department of Electrical-Electronics Engineeringen_US
dc.identifier.volume51
dc.identifier.issue7
dc.identifier.startpage887
dc.identifier.endpage895
dc.peerreviewedYesen_US
dc.publicationstatusPublisheden_US
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanıen_US
dc.contributor.institutionauthorKaraman, Mustafaen_US
dc.relation.indexWOSen_US
dc.relation.indexScopusen_US
dc.relation.indexPubMeden_US
dc.relation.indexScience Citation Index Expanded (SCI-EXPANDED)en_US
dc.description.qualityQ2
dc.description.wosidWOS:000222678000018
dc.description.pubmedidPMID:15301009


Bu öğenin dosyaları:

Thumbnail

Bu öğe aşağıdaki koleksiyon(lar)da görünmektedir.

Basit öğe kaydını göster